Analytical Instruments
HSM-DOG200 Dissolved Oxygen Meter
Panel-mount, 96 × 96

HSM-DOG200 technical data
Measurement
- Measuring range
- 0.00–19.99 mg/L; 0.0–199.9 %
- Resolution
- 0.01 mg/L; 0.01 %
- Accuracy
- ±1.5 % FS
- Control range
- 0.00–19.99 mg/L; 0.0–199.9 %
- Temperature compensation
- 0–65 °C
Electrical
- Supply voltage
- 230 V AC ±10 %, 50/60 Hz
- Output
- Isolated 4–20 mA
- Output load
- Max. 500 Ω
- Control output
- ON/OFF relay contacts
- Relay rating
- 5 A at 230 V AC; 10 A at 115 V AC
- Isolation
- 500 V DC minimum to earth
Mechanical and environment
- Dimensions
- 96 × 96 × 115 mm
- Panel cut-out
- 91 × 91 mm
- Weight
- 1 kg
- Ambient temperature
- 5–55 °C
- Relative humidity
- ≤ 80 %
- Magnetic environment
- No strong field beyond the earth's
Values are for the standard configuration. Ranges, threads, probe lengths and output types outside this table are available on request — the datasheet revision for a custom build is issued with the quotation.
Request the PDF datasheet →Analytical Instruments
HSM-DOG200Dissolved Oxygen Meter
Panel-mount, 96 × 96
Panel-mount dissolved oxygen transmitter reading 0–19.99 mg/L or 0–199.9 %, with automatic temperature compensation, isolated 4–20 mA output and dual relay control. 96 × 96 mm DIN case, 230 V AC supply.
Engineering reply · Evaluation samples available · Custom ranges built to order
- Reads 0–19.99 mg/L or 0–199.9 % saturation
- Automatic temperature compensation 0–65 °C
- Isolated 4–20 mA output
- Two ON/OFF relay outputs
- Standard 96 × 96 mm panel cut-out
- Ten-year parameter retention without backup battery
How it works
The DOG200 drives a polarographic dissolved oxygen electrode and displays the result in mg/L or percent saturation. It handles the compensation aeration processes actually need: temperature compensation from 0 to 65 °C, and an isolated 4–20 mA output so the signal reaching the PLC does not carry the plant's earth noise. Two relays with adjustable set-points can run a blower or an alarm directly. The 96 × 96 mm cut-out is the standard panel instrument size, which means it drops into an existing cut-out.
Typical applications
- Activated sludge aeration control
- Aquaculture and fish-farm oxygen monitoring
- Boiler feedwater deaeration
- Beverage and brewing process water
Not certain HSM-DOG200 is the right unit?
Send the medium, range, process connection and the signal your control system accepts. We will confirm the model — or tell you which one fits better — before you commit to an order.
Same family
Other analytical instruments

Dissolved Oxygen Electrode
Steam-sterilisable, 130 °C
Polarographic dissolved oxygen electrode that withstands 130 °C steam sterilisation, with a self-pressure-balancing design and an imported BJC membrane. 0.01–20 ppm range, 10 s response, 12 mm insertion diameter in 120, 280 or 450 mm lengths.
View specifications
pH / ORP Controller
Two analogue outputs, RS485
pH and ORP controller accepting a range of analogue electrodes through a built-in A/D stage. −2 to 16 pH, −2000 to 2000 mV, ±1 % FS, two 4–20 mA outputs, five relays, Modbus RTU, ABS housing, IP65.
View specifications